Cleaning Butterfly iQ

Screen_Shot_2019-05-06_at_2.33.13_PM.png

To clean the probe: 

  1. After each use of the probe, use one of the recommended liquid saturated wipes (Super Sani-Cloth® Germicidal Disposable Wipes by PDI, Inc., Super Sani-Cloth® AF3 Disposable Wipes by PDI, Inc., or a lint-free cloth moistened with water) to remove ultrasound transmission gel from the probe.
  2. Disconnect the probe from the mobile device.
  3. Wipe the probe, strain relief, cable, and connector with one of the recommended liquid saturated wipes for one (1) minute and until visibly clean.
  4. Change the wipes as necessary and repeat the above step until the probe is visibly clean.
  5. To dry the probe, use a soft cloth, blot the lens dry. Do not wipe the lens. Dry the rest of the probe, cable, strain relief, and connector.
  6. Visually inspect the probe in a well-lit area to ensure all surfaces are clean. If the probe is not clean, repeat the cleaning steps above.
  7. Dispose of cleaning material in accordance with all applicable regulations.