Disinfecting Butterfly iQ

After cleaning the probe, you must disinfect the probe.

It is recommended that you use Super Sani-Cloth® Germicidal Disposable Wipes by PDI, Inc or bleach (0.6% Sodium Hypochlorite) and clean non-linting wipes.

Screen_Shot_2019-05-06_at_2.49.09_PM.png

To disinfect the probe using Super Sani-Cloth® Germicidal Disposable Wipes by PDI:

  1. Wipe the probe, cable, strain relief, and connector with a Super Sani-Cloth® Germicidal Disposable Wipe. Use additional fresh wipes as needed.
  2. Make sure the treated surface remains visibly wet for a minimum of two (2) minutes paying attention to seams, gaps, gasket material, and recessed areas.
  3. Use additional fresh wipes as needed to ensure continuous two (2) minutes of contact time.
  4. Allow to air dry.
  5. Once clean and disinfected, visually inspect the probe, strain relief, cable, and connector for signs of damage or wear.

To disinfect the probe using bleach (0.6% Sodium Hypochlorite) and clean non-linting wipes:

  1. Wipe the probe, cable, strain relief, and connector using a clean non-linting wipe wetted (damp but not dripping) with bleach (0.6%). Use additional fresh wipes as needed.
  2. Make sure the treated surface remains visibly wet for a minimum of ten (10) minutes paying attention to seams, gaps, gasket material, and recessed areas.
  3. Use additional fresh wipes as needed to ensure continuous ten (10) minutes of contact time.
  4. Allow to air dry.
  5. Once clean and disinfected, visually inspect the probe, strain relief, cable, and connector for signs of damage or wear.

Disinfecting the Mobile Device

It may be necessary to disinfect your mobile device after use. For details, refer to your mobile device’s policy and website for support.

 

 

 

 

 

 